Reference Case 1 (Pelvis)
Plan a typical treatment using the attached field set-up.
Collimator settings are indicated in addition to field blocking
illustrated in the lateral and PA diagrams. Use a 60 ° (or the wedge
most apt to be used) for this set-up and treatment. The set-up may be isocentric or fixed SSD based on your institution’s
treatment technique.
Implement a weight ratio of 2:1:1 for the posterior and
lateral fields respectively. Create two treatment plans with the first photon
energy in the range of 4MV-10MV and the second photon energy greater than 10MV,
if available. Please calculate the monitor units or time set to deliver
1.80 Gy at the beam’s intersection.
Create the plans as if they were actual patients.
Be sure to include the independent monitor set calculation you normally
perform.
Calculations:
1.) Radiation Therapy Unit:
__________________________________________________________
Treatment
distance: SSD _________ cm; SAD _________ cm.
Low energy beam:
_______ MV; Wedge Angle ________; Transmission _________.
High energy beam:
_______ MV; Wedge Angle ________; Transmission _________.
Does the wedge
transmission vary with field size? YES / NO
(circle one)
Does the wedge
transmission vary with depth? YES / NO (circle
one)
Attach copies of
the transmission data
2.) Equation with variables used to
calculate required monitor set (mu or min).
Post. monitor set =
Lat. monitor set
=
Variables
defined:
3.) Please list the actual monitor set
used
Post. monitor set =
Left Lat. monitor
set =
Right Lat.
monitor set =
![]() |
![]() |
Pelvis |